5.0 PRE-OPERATIONAL FUNCTION CHECK
After reading the basic operation (Section 4.0), perform all the steps in this procedure.
1. Plug in the hand pendant cable-connector into the receptacle located at the head
end column.
2. Plug the power cord into a hospital grade power outlet and flip the main on/off power
switch to on. Verify the switch illuminates; this indicates power is applied to the table.
3. In succession, press and hold each of the six buttons on the hand pendant and
confirm the operation of the following functions:
3.1. Height up
3.2. Height down
3.3. Trendelenburg
3.4. Reverse trendelenburg
3.5. Tilt left
3.6. Tilt right
3.7. Engage the rotation safety lock (switch to on) and the 180° rotation locks (blue
handle clockwise, indicator light illuminates).
3.8. Release the rotation safety and 180° rotation locks.
3.9. Confirm head end and foot end crossbars are unlocked by rotating each one
complete revolution.
3.10. Re-engage both the rotation safety and 180° rotation locks.
3.11. Confirm head end and foot end crossbars are locked; both crossbars firmly
resist rotation.
3.12. Using the hand control, press and hold left lateral roll (TILT) button to end of
travel. Confirm the foot end crossbar rotates freely.
3.13. Using the hand pendant, press and hold right lateral roll (TILT) button to end of
travel. Confirm the foot end crossbar rotates freely.
Report any malfunction to your Mizuho OSI representative immediately.
MIZUHOSI 2009
23
NW0498 Rev E