8.0 TROUBLESHOOTING

8.1 Electrical System

In the advent of a base malfunction, the first item to investigate is the input power. Be
certain that the power cord is plugged into a live electrical outlet and the power switch is
switched on. A green "Power On" light illuminates indicates this.
Before removing any sheet metal covers the base must be turned off and
unplugged. Dangerous high voltage may be present in the circuitry under the
covers. Only trained technicians should perform this task.
SYMPTOM
No column function
either end.
No column function
Head end
No column function
Foot end
Rotation friction control
won't stop Rotation-Head end
Or Foot End.
Caster won't lock/unlock
MIZUHOSI 2009
WARNING:
INSPECT
Check power.
Check that circuit breaker (light is
green).
Check hand pendant module and
cord.
Check electrical connector at
base of head or foot end Column.
Check hand pendant.
Check electrical connector at
base of head end column.
Check hand pendant module and
cord. Check electrical connector
at the base of foot end column.
Tighten lever to max. Check
brake band.
Check caster for damage.
27
ACTION
Plug in.
Reset breaker, turn on.
Plug in connector.
Plug in connector.
Plug in connector.
Adjust brake tension.
Adjust handle to proper range.
Replace if necessary.
Replace caster.
NW0498 Rev E