Verify that the T-pin is inserted through the H-frame, passes completely through
the table mounting tube, and through the opposite side of the H-frame and the T-
pin drop lock is in sight and pivots freely.
4. To couple the head end of the table top, slide the H-frame and crossbar away from
the foot end as necessary to accommodate the length of the table top. Align the
holes in the head end H-frame and the table coupling device. Push the T-pin
completely through both sides of the H-frame and the coupling device making
certain that the drop lock is in sight and pivots freely. The head end and the foot end
of the table top are normally coupled at the same hole in the H-frame. The only
exception is when extreme trendelenburg is needed.
Failure to follow the above-prescribed procedures regarding securing the H-frame
to the crossbars and T-pin to the H-frame may result in the patient being dropped.
Prior to rotation, verify that all table coupling devices and H-frames are properly
installed and the T-pins pass completely through the H-frames and crossbars
with the drop lock in sight and pivots freely. Two (2) T-pins and drop locks should
be visible and checked at each end of the table.
4.6 Patient Transfer
Turn the power switch to on position. Prior to patient transfer, confirm that the desired
table top is secured by the H-frames and T-pins. Refer to specific table top coupling
procedure on page 16 of this Owners Manual for detailed information on attaching the
table tops. All three indicator lights should now be illuminated. If the tilt drive indicator
light is not lit, see synchronizing the lateral tilt function procedure on page 13.
If all three lights are not illuminated, do not use the base. Verify that the rotation
safety lock switch is in the on position and the 180° rotation lock lever is in the
locked position.
4.7 Table Top Coupling Procedure With Top In Place
When one table top is already coupled to the 5803 Advanced Control Base and you are
coupling a second top in order to change tops or facilitate a 180° tabletop rotation, the
slide mechanism of the head end sliding assembly will be held in a fixed position due to
the fixed length of the first top. This requires a different procedure to couple the second
top.
MIZUHOSI 2009
WARNING:
WARNING:
WARNING:
WARNING:
19