Troubleshooting - Dentsply Sirona CEREC AF Notice D'utilisation

Masquer les pouces Voir aussi pour CEREC AF:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10 PC Diagnostic Tool
10.3 Troubleshooting
80
Operating Instructions (valid for Canada) CEREC AF, CEREC AF Connect
➢ Check the result in the "Windows Stress Test" window.
Tip: You may have to rearrange the windows to be able to see the
"Windows Stress Test" window.
Test result: Pass
No errors were found on the PC-specific hardware. Replacing the PC
component or the PC is not advisable.
Test result: Fail
An error was found on the PC-specific hardware. Replacing the PC
component or the PC may be required.
An image restore is not advisable.
1. Check the "Windows Stress Test" window for those
components that did not pass the test.
2. Carry out the appropriate steps in the chapter entitled
Troubleshooting [ → 80] .
Ending the "Sirona Windows Diagnostics" test
➢ Click the "Stop" button.
 The individual test windows close.
 The PC is switched off.
10.3

Troubleshooting

The recommended procedure is as follows. If troubleshooting and/or a
component replacement cannot eliminate the fault, replace the PC.
In this case, specify which test was defective when returning the defective
PC.
Dentsply Sirona
66 37 487 D3652
D3652.201.01.14.02 10.2017

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Cerec af connectCerec aiCerec ai connect

Table des Matières