4.4.7 Output Current

Within this option the measuring range can be defined corresponding to the output current
of 4-20 mA.
The beginning of the
-
inverted signal, e.g. 0.5 to 6m corresponds to 20-4 mA.
The engineering units and decimal point selected will be valid within this option.
-
The basic setting of the span and offset can be checked within the test menu.
-
CURRENT
RELAY
In case of signal loss the device will emit an error signal of 22mA.
-
The diagram below shows the type of relationship between the 4-20mA ouptut and the
-
associated measuring range.
CALIBRATION MENU
measuring range might be larger than the end creating an
ENTER
4=00.00
4=02.00
Enter the end of the measuring range
mA
20
4
2
Enter the beginning of the
0......9
measuring range
ENTER
20=00.00
ENTER
20=12.00
12
46
0......9
m
8175