K-Factor; Output Current - Burkert 8025 Notice D'utilisation

Transmetteur de debit
Masquer les pouces Voir aussi pour 8025:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4 OPERATION

4.3.3 K-Factor

The K-Factor of the fitting is entered in this menu (see manual of fitting S020/1500/1501).
The "Teach in" function allows to practically detemine the application specific K-Factor.
The user only needs to run a known quantity through his system.
Example: In order to determine a quantity the most accurately possible, the user shall fill
a tank of 100 liters. When the message "TEACH YES" appears, he presses the Enter key to
start the measuring procedure. The message "FILL END" (end of filling) will appear. He then
switches on a pump or opens a valve. As soon as his tank is full, he switches off the pump
or closes the valve. Pressing Enter stops the measurement. The user will then be asked
to enter the quantity (100 liters). The calculated K-Factor is displayed after confimation.
Note: The device uses the K-Factor entered or determined at last.
ENTER
K-FACTOR
CURRENT
CURRENT

4.3.4 Output current

The measuring range of the flow, corresponding to the 4...20 mA output current is entered
here. E.g. 0 to 180 l/min corresponds to 4...20 mA. The beginning of the measuring range
can be larger than the end of it, i.e. 0 to 180 l/min corresponds to 20...4 mA (inverted output
signal).
The settings (unit and decimal place), as selected for the flow indication will apply.
8025 - 8025RR
FLOW TRANSMITTER 8025
ENTER
TEACH NO
ENTER
ENTER
TEACH Y
Start of measurement.
ENTER
Indication of calculated
K-Factor.
E-25-
K = =00.00
Entering of K-Factor as
specified in the charts.
0......9
K = 117.60
ENTER
FILL END
Stop of
measurement.
Entering the measured
quantity. Same unit as for flow.
ENTER
K = 79.30
0000.0 L
0......9
0100.0 L

Publicité

Table des Matières
loading

Table des Matières