4 OPERATION

4.3.3 K-Factor

The K-factor of the fitting (refer to S030 reference manual) is entered in this menu .
The "Teach in" function allows to practically detemine the application specific K-factor.
The user only needs to run a known quantity through his system.
Example: In order to determine a quantity the most accurately possible, the user shall fill
a tank of 100 litres. When the message "TEACH YES" appears, he presses the Enter key to
start the measuring procedure. The message "FILL END" (end of filling) will appear. He then
switches on a pump or opens a valve. As soon as his tank is full, he switches off the pump
or closes the valve. Pressing Enter stops the measurement. The user will then be asked
to enter the quantity (100 litres). The calculated K-factor is displayed after validation.
Note: The device uses the K-factor entered or determined at last.
ENTER
K-FACTOR
CURRENT
CURRENT

4.3.4 Output current

The measuring range of the flow, corresponding to the 4...20 mA output current is entered
here. E.g. 0 to 180 l/min corresponds to 4...20 mA. The beginning of the measuring range
can be larger than the end of it, i.e. 0 to 180 l/min corresponds to 20...4 mA (inverted output
signal).
The settings (unit and decimal place), as selected for the flow indication will apply.
!
In case of electronic internal failure, the current output is set to 22 mA.
FLOW TRANSMITTER 8035
ENTER
TEACH NO
ENTER
ENTER
TEACH Y
Start of measurement.
ENTER
Indication of calculated
K-factor.
E-18-
K = =00.00
Entering of K-factor as
specified in the charts.
0......9
K = 117.60
ENTER
FILL END
Stop of
measurement.
Entering the measured
quantity. Same unit as for flow.
ENTER
K = 79.30
0000.0 L
0......9
0100.0 L
8035