Offset Compensation; Span Compensation - Burkert 8175 Manuel Utilisateur

Transmetteur de niveau à ultrasons
Masquer les pouces Voir aussi pour 8175:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TEST MENU

4.5.1 Offset Compensation

Within this option the user has the possibility of correcting the basic setting of 4 mA
generated by the transmitter. The transmitter generates a value of 4mA by pressing
ENTER
when «OFFSET» is displayed within the main test menu.
Measure the generated current with an ammeter. If the displayed value is incorrect it can be
corrected by entering the measured value on the ammeter.
OFFSET
ENTER
OF= 04.00
Enter the measured value
0......9
SPAN
ENTER
OF= 04.02

4.5.2 Span Compensation

Span compensation provides the option of changing the basic setting of 20 mA. The
procedure is identical to that of the offset compensation above. The transmitter generates
is pressed when «SPAN « is displayed within the main test menu.
20mA if the key
ENTER
Measure the generated current with an ammeter. If the displayed value is incorrect it can be
corrected by entering the measured value on the ammeter.
SPAN
ENTER
SP= 20.00
Enter the measured value
0......9
T° ADJUST
ENTER
SP= 19.96
51