4 OPERATION

4.3.2 Temperature units

The temperature can be displayed in °Celsius or in °Fahrenheit.
ENTER
T° UNIT
CURRENT
ENTER

4.3.3 Output Current

Enter the measuring range corresponding to the 4...20 mA output. First select the unit pH
or T°current then, enter the limit values . E.g. 2 to 12 pH corresponding to 4...20 mA. The
beginning of the measuring range might be larger than the end of it, e.g. 2 to 12 pH corre-
sponds to 20...4 mA (inverted output signal).
CURRENT
ENTER
RELAY 1
ENTER
The minimum measuring range is 0,5 pH. If the beginning of the measu-
ring range equals the end of it, there will be no display of the current
value in the operation mode display (§4.2)
PH TRANSMITTER 8205
°CELSIUS
°FAHRENH
mA...PH
ENTER
mA...°C
Enter the
beginning of
the measuring
range
End of the
measuring
range
E-28-
4=00.00
0......9
4=02.00
ENTER
20=00.00
0......9
20=12.00