4.5 TEST MENU
ENTER
PRESS
0......9
The following parameters are selected and set within this menu:
OFFSET
SPAN
T° ADJUST
SIGNAL
0......9
SIMUL
RESET
END
The following sections explain how to change and investigate the parameter values within
the test menu above.
TEST MENU
SIMULTANEOUSLY FOR 5 SECONDS
Function
Zero point compensation (4 mA).
Span compensation (20 mA).
Temperature correction +/- 15°C .
Display of the returned signal strength.
Input of the level, volume or temperature to be si-
mulated. The current and relay outputs will react
according to this value.
Reset of the device to the factory settings and
resetting of the echo filtering table (see 4.4.5).
Return to the main menu and store the new para-
meters. If one of the two values for OFFSET and
SPAN is inappropriate, the device will automa-
tically return to the «OFFSET» parameter and new
values must be entered.
50
Section
4.5.1
4.5.2
4.5.3
4.5.4
4.5.5
4.5.6