Pump and Electrolyte
Container
Every day
Cabinet
Every day
Every week
Once a year
Exchange of Batteries
Movipol-3
Instruction Manual
3. Maintenance
The electrolyte container must be emptied and rinsed with water
after use. The pump system must also be rinsed out. To do this,
circulate water from the electrolyte container through the pump
system.
Electrolyte spilled on the front plate or other parts of the cabinet
should be wiped off.
Stains of metal oxide can be removed by means of the cleansers
and sealants used for cars and fiberglass boats.
Treat the cabinet surface with a surface sealant once a year in
order to preserve it.
For general cleaning purposes use a cloth and a weak detergent.
The batteries supplied will last for several years provided they are
used in a correct way. In case of a full discharge, however, the
batteries may be damaged and will have to be exchanged.
A battery kit (9 pcs.) can be ordered from your Struers
representative; see section CONSUMABLES AND ACCESSORIES.
Connections must be done correctly (+ pole to ÷pole); see fig. 11.
Never leave damaged batteries in the apparatus for an extended period.
Follow the procedure for changing the battery fuse; see fig. 11.
Remove the 3 rails (fig. 11.3, 6 screws (fig. 11.4)).
Disconnect the old batteries and remove them.
Place the new batteries in exactly the same position as the old
ones.
Replace the 3 rails and the cabinet.
Charge the batteries; see section INSTALLATION.
IMPORTANT