Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Depending on the confi guration, the screen will
now display
- the active port (right or left) and
- the percentage of light going to the port (100%,
80%, 50%).
Optional Leica DMI 6000 B:
The bottom port selection function can be as-
signed to one of the variable function buttons on
the stand or the SmartMove.
The top port can only be selected manually.
Select ports via
fi xed function buttons on stand (side ports)
variable function buttons on stand and
SmartMove (bottom port)
manual action (top port)

8.8 Eyepieces

Note:
The eyepiece's aperture protector must be re-
moved or folded back, during microscopy while
wearing eyeglasses. We recommend removing
bifocals and spectacles with progressive-addi-
tion lenses when using the microscope.
• For the adjustable tubes with documentation
output, choose the 100% VIS position.
Eyepieces with inlaid reticle
• Focus the reticle by adjusting the eyelens.
• Focus on the object through this eyepiece.
• Then, close that eye and focus on the object
by adjusting only the second ocular.
Correction for Vision Problems
• With your right eye, look through the right eye-
piece and bring the specimen into sharp focus.
• Then, with your left eye, view the same speci-
men and rotate the left eyepiece tube until
the object is brought into sharp focus. Do not
change the Z position in the process!
8. Operation
97

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Dmi4000 bDmi6000 b

Table des Matières