3. Safety Notes
3.2 Electrical Safety
General Specifi cations
Leica CTR4000, CTR5000, CTR5500, CTR6000,
CTR6500, CTR7000, CTR6500 HS, CTR7000 HS
Electronics Boxes
For indoor use only.
Supply voltage:
Frequency:
Power input:
Fuses:
Ambient temperature:
Relative humidity:
Over voltage category:
Pollution degree:
Microscope
For indoor use only.
Supply voltage:
Frequency:
Power input:
Fuses:
Ambient temperature:
Relative humidity:
Over voltage category:
Pollution degree:
10
90–250 V~
50–60 Hz
max. 290 VA
T6.3 A
(IEC 60127-2/3)
15–35°C
max. 80% to 30°C
II
2
90–250 V~
50–60 Hz
See CTR4000–7000 HS
See CTR4000–7000 HS
15–35°C
max. 80% to 30°C
II
2
ebq 100 supply unit*
For indoor use only.
Supply voltage:
Frequency:
Power input:
Fuses:
Ambient temperature:
Relative humidity:
Over voltage category:
Pollution degree:
(see enclosed manual)
Leica EL6000*
For indoor use only.
Supply voltage:
Frequency:
Power input:
Fuses:
Ambient temperature:
Relative humidity:
Overvoltage category:
Pollution degree:
(see enclosed manual)
90–250 V~
50–60 Hz
see CTR4000–7000
see CTR4000–7000
15–35°C
max. 80% to 30°C
II
2
100–240 VAC
50–60 Hz
max. 200 VA
5x20, 2.5 A, slow,
breaking capacity H
0°–40°C
10–90%
non-condensing
II
2