8. Operation
Select objectives via
variable function buttons on stand and
SmartMove
Leica Application Suite (LAS) Software
Manual selection
Changing the operating modes
„dry" ( DRY) and „immersion" ( IMM)
Each objective is assigned to a specifi c objective
category:
1) Dry objectives (DRY)
2) Immersion objectives (IMM)
Note:
It is possible to use objectives for both operating
modes.
The mode can be assigned in the Leica Applica-
tion Suite (LAS).
Changing the operating mode
• First, select the operating mode (Imm or Dry)
using the function buttons.
The operating mode may also be selected in
the Leica Application Suite (LAS).
• The objective turret is lowered to its bottom
stop. This is to permit the application of the
immersion liquid when changing from a dry to
an immersion objective. It also permits the re-
moval of the liquid when changing to dry mode.
The current objective remains in the beam
path.
100
• Next, press the but on for the objective you in-
tend to use.
Note:
If the Imm or Dry operating mode buttons are
pressed accidentally, the original mode can be
restored by pressing the appropriate button.
Change operating mode via
variable function buttons on stand and
SmartMove
Leica Application Suite (LAS) Software
Note:
When replacing objectives, you must perform
a teach-in for the new objectives in the Leica
Application Suite (LAS). A parfocality teach-in
should also be performed.
Note:
For lockable immersion objectives lock these by
pushing the front part upwards until it stops (ap-
prox. 2 mm). Then, after a gentle turning motion
to the right, the objective is locked.
For objectives with corrective mounts turn the
knurl to adjust the objective to the thickness of
the cover glass.