Motorized 3-plate or Scanning Stages
3-plate stages and scanning stages: after install-
ing the stage, connect the included stage cable
(for motorized stages) fi rst to the socket on the
stage, then to the CRT6000, CTR6500 or CTR7000
box. The correct place on the box is called „XY
Stage".
A variety of inserts (including heating ones) are
available for the normal 3-plate and scanning
stages. Install these inserts diagonally from
above into the corner with the spring clips. The
insert will click into place when seated properly.
Fig. 29 a, b Mounting screws for 3-plate cross-stage
a
b
!
Caution:
Press the spring clip into place only from the
side.
Do not press the insert onto the spring clips di-
agonally from above, as the insert will not be
aligned parallel to the stage and may be bent in
the process.
6. Assembly