Leica Microsystems DMI3000 B Mode D'emploi page 51

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

• Insert the lamp mount, with the burner in-
stalled, into the lamp housing and tighten it
with the screws (63.8).
• Test the adjustment of the collector (63.2):
Do not touch the power supply while perform-
ing these actions. When closing the lamp
housing, ensure that the pins of the contact
plug engage in their sockets (63.9).
Tighten the screws of the cover and press the
contact plug home.
• Place the lamp housing in the incident light
lamp housing mount (Fig. 53) and fasten it with
the clamping screw on the side.
• Connect the lamp housing to the external pow-
er supply (62.1).
Caution!
The burner must be adjusted immediately af-
ter lighting.
Leica EL6000
Caution!
When using the compact light source Leica
EL6000, it is essential to observe the safety
information in the separate instructions.
6. Assembly