8.2 Disinfecting

8.3 Sterilizing

9. FUNCTION CHECK

130 •
seca mBCA 555/554seca mBCA 552
1. Use an alcohol-based disinfectant (e.g. 70 % ethanol).
2. Follow the instructions on the disinfectant.
3. Disinfect the device:
► Moisten a soft cloth with disinfectant and wipe down the device with
it.
► Comply with the intervals, see table.
Interval
Before every
measurement
After every
measurement
If required
This device may not be sterilised.
► Perform a function check prior to each use.
A complete function check includes:
• Visual inspection for mechanical damage
• Checking the alignment of the device
• Visual and function check of the display elements
• Function check of all the controls shown in the section entitled
"Overview"
• Function check of optional accessories
If you notice any faults or deviations during the function check, first try to
resolve the error with the aid of the section entitled "Troubleshooting" in this
document.
CAUTION!
Personal injury
If you notice any faults or deviations during the function check which
cannot be resolved with the aid of the section entitled
"Troubleshooting" in this document, you may not use the device.
► Have the device repaired by seca Service or by an authorized ser-
vice partner.
► Follow the section entitled "Servicing" in this document.
Component
• Weighing platform
• BIA standing aid with hand electrodes
• Weighing platform
• BIA standing aid with hand electrodes
• Multifunctional display
• Column (devices with measuring rod)
• Measuring head (devices with measuring
rod)