11.2 Non-calibrated scales
11.3 Bioimpedance measurement
138 •
seca mBCA 555/554seca mBCA 552
5. Read off the calibration counter.
The value must match the number indicated on the calibration counter
seal (figure shows example values).
Both numbers have to match for the verification to be valid. If the sticker and
the verification counter do not match, the scale must be verified. Please con-
tact your service partner or seca Service. Once the scale has been verified, a
new, updated verification counter sticker is used to identify the status of the
verification counter. The person authorized to perform the verification secures
this sticker with an additional seal. The verification counter sticker can be
ordered from seca Service.
The product must be set up carefully and serviced regularly. Depending on
how frequently the product is used, we recommended servicing at intervals of
3 to 5 years.
CAUTION!
Incorrect measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service or
by an authorized service partner.
► You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
The bioimpedance analysis (BIA) measuring technology must be checked
every two years. We recommend servicing the whole device as part of this
check.
NOTICE!
Incorrect measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service or
by an authorized service partner.
► You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.