Height Measurement, Ultrasonic - Seca mBCA 555 Mode D'emploi

Masquer les pouces Voir aussi pour mBCA 555:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Fault/error code
Autohold
function cannot
be deactivated
001-272XX-XXX
to
008-272XX-XXX
020-272XX-XXX
to
023-272XX-XXX
010-272XX-XXX
013-272XX-XXX
019-272XX-XXX

10.3 Height measurement, ultrasonic

Fault/error code
symbol appears
during calibration
Ultrasound measuring
head status LED does not
light up
132 •
seca mBCA 555/554seca mBCA 552
Cause
Device is connected to a network
(intended behavior): Autohold function
is activated automatically.
Load cell or weight calculation module
defective
One corner of the scale has been
loaded excessively.
The scale has been switched on with
too high a load.
The scale was caused to oscillate and
was unable to determine the zero point.
Ambient temperature too high or too
low.
Cause
Calibration failed
Device is in an undefined state
following implausible inputs.
Wiring in the ultrasound measuring
head incorrect
The status LED is defective.
Remedy
If necessary, disconnect device from
network
Inform seca Service
• Distribute weight evenly
• Disconnect the plug-in power supply unit
from the socket
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Inform seca Service
• Remove the weight from the weighing
platform
• Disconnect the plug-in power supply unit
from the socket
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Disconnect the plug-in power supply unit
from the socket
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Observe ambient conditions for operation,
transport, and storage ➔
technical data, page 142
Remedy
• Make sure that no objects or people are in
the immediate vicinity of the device during
calibration
• Make sure that the supplied calibration
rod was used
• Make sure that the calibration rod is posi-
tioned centrally on the foot silhouettes of
the weighing platform
• Take power supply unit out of the socket
• Wait approx. 1 minute
• Plug the power supply unit into the
socket; the device and the multifunctional
display switch on automatically
Wire the ultrasound measuring head as
described in the corresponding assembly
instructions
Inform seca Service