Fault/error code
02-288XX-XXX
052-280XX-XXX
420-280XX-XXX
10.6 Display message traffic light system
Symbol
10.7 Error codes
136 •
seca mBCA 555/554seca mBCA 552
Cause
No data transmission possible
Internal device communication error
No connection to the server
Green:
Action successful, e.g. data submitted to EMR System or
seca analytics 125
Yellow:
Incorrect operation or malfunction, can be remedied by the user with the aid
of the error tables in these instructions for use ➔
page
131.
Red:
Device error that cannot be remedied by the user. Inform seca Service.
420 - 288BA - 555
Check network and workflow settings:
• Check network settings ➔
network functions, page 121
• Check workflow settings:
seca connect 103
use
• Disconnect the plug-in power supply unit
from the socket
• Wait approx. 1 minute
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Inform seca Service
• Check LAN cable
• Check network settings ➔
network functions, page 121
Description
software
Model number; here:
seca mBCA 555/554
Firmware version of the assembly concerned; here:
Index "A"
Hardware version of the assembly concerned; here:
Index "B"
Assembly concerned; here: 288 = interface module
Error number; here: 420 = no connection to the server
Remedy
Setting up
system instructions for
Setting up
Troubleshooting,
scale