Disinfecting; Sterilizing; Function Check - Seca 635 Mode D'emploi

Masquer les pouces Voir aussi pour 635:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 103

8.2 Disinfecting

8.3 Sterilizing

9. FUNCTION CHECK

ATTENTION!
Damage to device
The viewing windows of scales and displays are
made of acrylic glass. Acrylic glass windows
can crack or become cloudy if unsuitable disin-
fectants are used on them.
– Only use disinfectants suitable for sensitive
surfaces. Suitable disinfectants are available
from specialist dealers.
1. Follow the instructions on the disinfectant.
2. Disinfect the device at regular intervals using a soft
cloth dampened with a suitable disinfectant.
3. Please take note of the following terms:
Interval
Prior to every measurement
with direct skin contact
After every measurement
with direct skin contact
If required
The sterilisation of this device is not permitted.
Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• Function check of all the controls shown in the
section entitled "Overview".
• a function check of the optional accessories
English
Component
Weighing platform
Weighing platform
• Display
• Membrane keyboard
Function check • 93

Publicité

Table des Matières
loading

Table des Matières