Data Transmission - Seca mBCA 555 Mode D'emploi

Masquer les pouces Voir aussi pour mBCA 555:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Fault
is displayed
In
seca analytics 125
software:
Results of bioimpedance
measurement deviate
significantly from expected
results
In
seca analytics 125
software:
Value of an analysis
parameter is shown in red
037-276XX-XXX

10.5 Data transmission

Fault/error code
The "Date of birth" dialog
window appears after
pressing the
key
Autoclear function: Entering
"0 seconds" is not accepted;
instead, the factory setting
(300 seconds) is suggested
Workflow LED does not
light up
Workflow LED lights up red
Cause
Bioimpedance measurement failed
Patient moved during the
measurement
Patient used different pairs of hand
electrodes on the left and right
Electrodes defective
Value outside of the normal range
determined for this analysis parameter
Ambient temperature too high or too
low.
Cause
Patient ID not scanned
ID scanning not set up
Patient does not yet have a file in the
EMR System
Device is connected to the
software
seca analytics 125
Implausible input; switching off the
function is not intended in the factory
No network connection
WiFi function deactivated
Workflow LED is defective
ID was
found in the EMR System or
not
the seca software
Device has
saved measuring
not
results to the clipboard.
The measuring results were
submitted to an EMR System or
seca software.
• Clear measurement procedure with
key
• Check patient's electrode contact
• Repeat bioimpedance measurement
Request the patient not to move during the
measurement and repeat the
measurement
Ensure that the patient uses the same
hand electrodes on both sides and repeat
the measurement
Inform seca service
• Repeat the measurement to rule out
measurement error
• If the value still lies outside of the normal
range, take this into account during
evaluation and further examinations
• Observe ambient conditions for operation,
transport, and storage ➔
technical data, page 142
• Press
• Scan patient ID
Check workflow settings for the device in
the
seca connect 103
• Enter date of birth and press
again
• Create patient file in the EMR System and
assign measurement
• Enter date of birth
• Press
• Accept factory setting
• Manually enter value between 1 and 3600
seconds
Set up network connection ➔
network functions, page 121
Activate WiFi function ➔
deactivating the WiFi function, page 123
Inform seca Service
Create the ID in the EMR System or the
seca software
Repeat measurement
not
• Repeat measurement
• Check WiFi connection
English
Remedy
General
Remedy
key
software
key
key
Setting up
Activating/
Troubleshooting • 135