7. HYGIENE TREATMENT
7.1 Cleaning
7.2 Disinfecting
WARNING!
Electric shock
The device is not de-energized when the on/off
key is pressed and the display goes out. Use of
fluids on the device may cause electric shock.
– Before each hygiene treatment, remove the
mains cable to de-energize the device.
– Ensure that no fluids penetrate the device.
CAUTION!
Damage to device
Inappropriate detergents and disinfectants may
damage the sensitive surfaces of the device.
– Do not use aggressive or abrasive cleaning
agents.
– Do not use white spirit or petroleum spirit.
– Only use disinfectants suitable for sensitive
surfaces. Suitable disinfectants are available
from specialist dealers.
Use a soft cloth dampened with mild soapsuds to
clean the surfaces of the device.
ATTENTION!
Damage to device
The viewing windows of scales and displays are
made of acrylic glass. Acrylic glass windows
can crack or become cloudy if unsuitable disin-
fectants are used on them.
– Only use disinfectants suitable for sensitive
surfaces. Suitable disinfectants are available
from specialist dealers.
1. Follow the instructions on the disinfectant.
2. Disinfect the device at regular intervals using a soft
cloth dampened with a suitable disinfectant.
English
Hygiene treatment • 75