10.0
10.0
Specifi cation
Mains voltage (incl. tolerance)
arranged by user / installer by inserting the correct fuse drawer
Mains frequencies (incl. tolerance)
Mains connection
Auxiliary-mains power socket (IEC 320)
Power rating
Blower air fl ow (free-fl ow)
Device air fl ow
Activation
Operating mode
Protective measures (Fuses,...)
Fuse (Breaking capacity H)
Acclimatisation rules
Maintenance
Designed freedom from maintenance
Corrective maintenance
Inspection
Ambient conditions
Ambient temperatures
Air humidity
Atmospheric pressure
Weight:
Dimensions (H x W x D )
Noise level
Interfaces
Class of protection
Applied Part
Main-fi lter
Canadian Classifi cation
Device Group
PNC
Risk Class
Description
UK
AtmoSafe WORLD: 100...230VAC ± 10% (total range 90...253VAC) can be
AtmoSafe EUROPE: 230 V ± 10 %
50...60 Hz +/-1%
Device inlet connector for non-heating apparatus
120 V / 9.7 A - 230 V / 6.3 A
Max. 400 W
1600 l/min
650 l/min electronically controlled
Automatic activation, front-panel button, signal at activation input,
optionally by foot switch
Suitable for continuous operation
The power pack is fi tted with fusible cutouts 5 x 20. The motor of the
blower is thermally protected.
100 V: T4A (125 V) / 120 V: T4A (125 V) / 230 V: T3,15A (250 V)
(Prior to start-up): 6 h acclimatisation after transport at low temperature
(Acc. to DIN 31051 Measures for maintaining the desired condition):
Filter change and change of single-use articles (by user).
The basic device is maintenance-free.
>10 Years (20,000h) (blower, bearings, rubber parts, etc. )
(Acc. to DIN 31051 Measures for restoring the desired condition):
Corrective maintenance by ATMOS or by customer service authorised
by ATMOS in accordance with the country-specifi c regulations
(in Germany DIN VDE 0751).
(Acc. to DIN 31051 Determination of current state):
No repeated safety-practical checks as understood by MPG§§22-24.
(German law on medical products)
(In operation and in transport / storage):
In operation:
+10 ... +40°C / in transport and storage: - 40 ... +70°C
5...95% (without condensation)
+700...1060 hPa
14 kg
H 210 mm x W 368 mm x D 410 mm
Max. 52 dB(A)@1 m (acc. to ISO 7779)
(Input and output) suction hose connection 7/8"(22 mm),
equipotential bonding, activation input, mains connection
I
CF
defi brillator-protected
ULPA, retention = 99.9999 % @ 0.01µm
General Hospital & Personal use
80FRN
3
PUMP, INFUSION
Issure Technical Specifi cations 01.04.2005
Page 44
F
43

Publicité

Table des Matières