10.0
10.0 Technical specifi cations
Voltage
Current input
Power cable
Power consumption
Operating time
Height adjustment
Seat depth
Seat height
Fuse
Protective earth conductor resistance
Earth leakage current
Ambient conditions
Transport/storage
Operation
Dimensions HxWxD
Weight
Max. weight of the patient
Protection class (EN 60601-1)
Degree of protection
Protection category
Classifi cation acc. to Annex IX
EEC directions 93/42/EEC
CE marking
Rules applied
UMDNS-Code
Reference-No.
Canadian Classifi cation
Device group
PNC
Risk Class
Description
UK
230 V~ ± 10 %
50/60 Hz
3.0 A
3 m
670 W (=power of column)
Short-time operation, 2 min max.
electrically-driven lift column,
operation by foot lever
47 cm
54 - 74 cm
T 3,1 A
0.1 Ω
0.5 mA
-20...+50°C
5...95 % humidity without condensing,
air pressure 700...1060 hPa
+10...+40°C
30...95 % humidity without condensing,
air pressure 700...1060 hPa
138 x 65 x 80 cm
90 kg
150 kg
I
Type B
IPX0
I
CE
EN 60601-1: 1990 + A1:1993 + A2:1995
EN 60601-1-2: 1993 (EMV / EMC)
10-794
503.0338.0
General Hospital & Personal use
80FRK
1
Chair, Examination and treatment
Issue of the Technical Specifi cations: 04.06.2010