3-way valve (only for centrifuges with 3-way valves)
To thoroughly rinse out the system, the 3-way valve can be switched from physiological saline solution (NaCl) to
distilled water (H
O).
2
ROTOLAVIT
NaCl
H O
2
ROTOLAVIT
10
Opening and closing the lid
10.1
Opening the lid
The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on "Emergency release".
•
Swing handle rail on the lid upwards. The
•
Open the lid.
10.2
Closing the lid
Do not bang the lid shut.
•
Place the lid and swing handle rail on the lid downward. The
11
Inserting and removing the rotor
E
•
A
•
B
C
D
Connections:
NaCl: Supply hose (∅ 7 mm) of the physiological saline
solution.
H
O: Supply hose (∅ 7 mm) of the distilled water.
2
ROTOLAVIT: Connecting hose to the device.
Positions of the 3-way valve:
NaCl: The system is connected to the physiological saline solution.
H
O: The system is connected to the distilled water.
2
LED will now go off.
Hold the rotor down into the recessed grip (A) and place it on the hub (C). Both the
driving plates (B) in the rotor must be located in the two grooves (D) of the hub
(C). The alignment (E) of the driving plates (B) is indicated on the rotor.
Removing the rotor: Hold the rotor by the recessed grip (A) and lift it off upwards.
LED will come on.