3.
Decanting.
Clean and DCNT is displayed.
PROG
CYC
ml
RPM
Clean
600 DCNT
4.
Shaking.
Clean and AGIT is displayed.
PROG
CYC
ml
RPM
Clean
0
5.
Decanting.
Clean and DCNT is displayed.
PROG
CYC
ml
RPM
Clean
600 DCNT
•
After completion or after termination of the cleaning program by pressing the
with the fixed braking level. The braking level is displayed.
Display with stationary rotor:
PROG
CYC
ml
RPM
t/ min:s
OPEN LID
•
Open the lid.
Display after completion of the cleaning program:
PROG
CYC
ml
RPM
t/ min:s
Clean proc. DONE
If the cleaning program is terminated with the
display appears:
PROG
CYC
ml
RPM
t/ min:s
Clean Program
The cleaning program must be started again.
•
Switch the unit OFF.
Filling the system:
•
Device without 3-way valve:
Take off the supply hose from the distilled water container and connect it again to the physiological saline
solution container.
Device with 3-way valve:
Turn the 3-way valve to the "NaCl" position so that the system is connected to the physiological saline solution;
see the chapter on the "3-way valve".
•
Switch the unit ON
Display:
PROG
CYC
ml
RPM
t/ min:s
Clean proc. DONE
If
OPEN LID
•
Hold a container with a capacity of at least 80 ml in front of the injection pipe in order to catch the distilled water.
Then press and hold the
If a different key than the
PROG
CYC
t/ min:s
ml
RPM
Press 3s SALINE
The system is filled with saline solution.
Display:
PROG
CYC
ml
RPM
t/ min:s
Refill process
56/130
t/ min:s
t/ min:s
AGIT
t/ min:s
key before the distilled water was decanted the following
STOP
is displayed open the lid.
key for 3 seconds.
SALINE
key is pressed the following display appears:
SALINE
key the run-out is performed
STOP