Query The Operating Hours, Centrifugation Runs And Cycle Counter; Request System Information; Immediate Display Of Centrifugation Data After Switch-On; Cooling (Only In Centrifuges With Cooling) - Hettich ROTINA 380 Mode D'emploi

Masquer les pouces Voir aussi pour ROTINA 380:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

30

Query the operating hours, centrifugation runs and cycle counter

The operating hours are divided up into internal and external operating hours.
Internal operating hours: Total time the device was switched on.
External operating hours: Total time of previous centrifugation runs.
The query is performed as follows when the rotor is at a standstill:

Keep the
key held down for 8 seconds.
PROG
After 8 seconds,  Machine Menu  appears in the display.

Press the
key as often as necessary until the -> Operating Time is displayed.
PROG

Press the
key. The external operating hours are displayed, e.g. OP Time ext = 0h25m.
START

Press the
key. The internal operating hours are displayed, e.g. OP Time int = 1h36m.
PROG

Press the
key. The number of all centrifugation runs is displayed, e.g. Number of Starts = 10.
PROG

Press the
key. The number of rotor cycles (centrifugation runs) of the current rotor code since the last time
PROG
the cycle counter was reset to "0" and the permitted number of rotor cycles are displayed, e.g. CYCLES = 5120
of 30000.

Press the
key. The number of all rotor cycles (centrifugation runs) of the current rotor code is displayed,
PROG
e.g. Rotor cycles total = 37490. This value cannot be set.

Press the
key twice to leave the menu, "Operating Time" or press the key
OPEN / STOP
to leave the "Machine Menu".
31

Request system information

Rotor information and the program version of the centrifuge as well as the frequency converter can be queried as
follows when the rotor is at a standstill:

Keep the
key held down for 8 seconds.
PROG
After 8 seconds,  Machine Menu  appears in the display.

Press the
key as often as necessary until the ->Info is displayed.
PROG

Press the
key. The rotorcode (rotor), the maximum rotor speed (Nmax) and the preset centrifuging radius
START
(R) of the last rotor that was recognized by the rotor recognition , e.g. Rotor 4: Nmax = 4500 R=184.
The last recognized rotor is marked with a star (). Use the adjusting knob
permitted in the centrifuge displayed.

Press the
key. The program version of the centrifuge is displayed, e.g. SW version = 01.00.
PROG

Press the
key. The program version of the frequency converter is displayed e.g. FC-SW-Version = 4.
PROG

Press the
key twice to leave the "Info" menu or press the key
OPEN / STOP
"Machine Menu".
32

Immediate display of centrifugation data after switch-on.


Switch on the power supply. (Switch setting ).

The first time the display changes optically, (inverse display) press any key and keep it held down. The
centrifugation data is displayed immediately.
33

Cooling (only in centrifuges with cooling)

The temperature set-point can be adjusted from -20°C to +40°C / -4°F to +104°F. On centrifuges with a
heating/cooling option, the temperature set-point is settable from -20°C to +90°C / -4°F to +194°F. The lowest
obtainable temperature is dependent on the rotor (see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and
accessories").
33.1

Standby-cooling

When the rotor is at a standstill and when the lid is closed, the centrifugal chamber is cooled down to the preselected
temperature if this is lower than 20°C / 68°F.
During the standby cooling, the preset temperature is displayed.
three times
OPEN / STOP
to have the information of the rotors
three times to leave the
OPEN / STOP
EN
63/165

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Rotina 380 r

Table des Matières