Push button [14] to store the treatment step. The treatment step number is automatically incremented;
Repeat the previous two steps until the end of your protocol has been reached;
Terminate your protocol with a treatment step with a treatment time of zero. Push button [14] to
confirm. The end of your protocol is confirmed with a beep.
Notes:
When two consecutive treatment steps contain AC current waveforms, the current intensity will be
maintained at the adjusted level when the protocol is executed.
When switching to or between DC current waveforms, the current intensity will be decayed to zero.
The next treatment step will have to be started manually by adjusting the current intensity.
5.5
Connection and disconnection reactions
Constant Current (CC) output characteristics may cause unpleasant connection and disconnection
reactions if the electrodes are not securely placed or lose contact with the skin. Make sure the output is
set to 0 mA when you apply or remove the electrodes. Preferably use Constant Voltage (CV) in case of
dynamic application techniques.
5.6
Electrolytic effects
Electrolysis occurs under the electrodes when current types with a DC component are applied. Because
the greatest concentration of electrolytic by-products caused by ion migration occur under the electrodes,
we recommend the use of the supplied sponges to keep the effects to a minimum. Make sure that the
sponges are kept well moistened and place the thick side of the sponge between the flexible rubber
electrodes and the patient.
5.7
Remote control
For the Endomed 482 a special remote control is available. The remote control can be connected to
connector [3-2]. The remote control has 2 intensity buttons for setting the intensity of the current channels
and an emergency stop.
6 Maintenance
6.1
Technical maintenance
Electrical safety of the apparatus depends on the electrical connection (via the power cord) to protective
earth. It is therefore necessary to have this connection checked annually. Further an annual overall check
is recommended. This may be done by your supplier, or by another agency, authorized by the
manufacturer. It is also recommended that a record of the service history is kept for all activities relating
to service and maintenance. In some countries this is even obligatory.
Maintenance and all repairs should only be carried out by an authorized agency. The manufacturer will
not be held responsible for the results of maintenance or repairs by unauthorized persons.
The service and/or maintenance must be carried out conform the procedure described in the service
manual of the apparatus.
Opening of the equipment by unauthorized agencies is not allowed and will terminate any claim to
warranty.
6.2
Cleaning of the apparatus
Switch off the apparatus and disconnect it from the mains supply. The apparatus can be cleaned with a
damp cloth. Use lukewarm water and a non-abrasive liquid household cleaner (no abrasive, no alcohol
content solution).
GB
14