Télécharger Imprimer la page

Enraf Nonius Endomed 482 Mode D'emploi page 35

Masquer les pouces Voir aussi pour Endomed 482:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 73
1 Introduction
1.1
General
The Endomed 482 is equipped with two completely identical multi functional electrotherapy channels. The
electrotherapy channels can be used in combination or totally independent. These combination
opportunities along with the largest possible quantity of available current waveforms make this apparatus
the ultimate solution for your professional activities.
The apparatus is only meant to be used by competent personnel in physiotherapy, rehabilitation or
adjacent areas.
1.2
Therapy possibilities
Electrotherapy with low frequency, medium frequency, TENS, High Voltage and Microcurrent currents.
Through the application of these currents it is amongst others possible to diminish muscle tone, manage
pain and increase the circulation and regeneration of tissue..
1.3
Finally
You have made a wise choice in selecting the Endomed 482. We are confident that your unit will continue
to give satisfaction over many years of use. Nevertheless, if you have any queries or suggestions, please
contact your authorised distributor.
2 Safety
2.1
Preliminary notes
It is important that you read these operating instructions carefully before using the Endomed 482. Please
make sure that these instructions are available to all personnel who operates the equipment.
Pay attention to the following before using the Endomed 482:
1. Keep yourself informed of the contra-indications (see chapter 3).
2. Never leave the patient unattended during treatment.
3. The apparatus may not be used in close proximity (i.e. less than 2 metres) to shortwave equipment.
4. The apparatus may not be used in so-called "wet rooms" (hydrotherapy rooms).
The manufacturer cannot be held responsible for the results of using this apparatus for any purposes
other than described in these operating instructions.
2.2
General
When the apparatus is switched on all vital functions are checked by the internal microprocessor. During
operation the apparatus the actual output of the current is measured continuously and compared to the
requested output. Whenever a faulty situation is detected the apparatus will switch off the current
immediately.
2.3
Exclusions
Electrotherapy is not intended for the following specific applications:
Intercranial treatment, like sleep therapy;
Transcordial treatments (applying electrodes at the thorax may increase the risk of heart fibrillation);
Cervical occipital treatments.
2.4
Current density
The particular standard for Therapeutic Electrical Stimulation Equipment is IEC 60601-2-10 (BS 5724 part
2.10). The maximum permissible Current Density is 2 mA R.M.S. per cm². To verify that this value is not
exceeded in electrotherapy, divide the current output in mA by the effective electrode area in cm²; a value
of less than two should result.
Note that placing the electrodes should be done carefully and solidly, and that the use of small electrodes
in combination with high intensities may cause skin irritations or burns.
GB
1

Publicité

loading