E Wound healing:
skin lesions due to circulation disorders;
post-operative wounds.
3.2
Absolute specific contra-indications
3.2.1
Electro therapy
pyrexia;
tumours;
tuberculosis;
localized inflammation;
thrombosis;
pregnancy;
pacemakers;
metal implants.
4 Installation
4.1
Connection
Mains supply connections must comply with the national requirements regarding medical rooms. The
equipment has a safety earth (ground) connection, and must be connected to an earthed (grounded)
wall socket.
Prior to connection of this apparatus to the mains supply, check that the voltage and frequency stated
on the type plate correspond with the available mains supply.
The mains adapter is a part of the supply circuit on which the device's safety partly depends. The
approvals for the Endomed 482 are only valid if used in combination with this type of adapter ENA
1524
It is not permitted to connect the Endomed 482 to another type of adapter than type ENA-1550.
4.2
Connection of the mains adapter
Connect the supplied mains adapter to the connector [3-1].
Connect the mains adapter to a wall socket. Lamp [6] indicates that the apparatus is connected to the
mains supply and stand-by.
4.3
Switching on and self test
Switch On the apparatus, using the On/Off switch [1]
Immediately after switching on, the apparatus carries out a self test.
At the end of the self-test a beep is heard. When an error is found an error code will appear on the
display. See section 7 for details.
4.4
Disconnecting from the mains
Switch Off the Endomed 482, using the mains switch [1].
Pull out the mains plug of the mains adapter from the wall socket.
4.5
Battery operation (optional)
The apparatus can be operated independent of the mains supply. To make this possible, first a battery
must be installed. This is done as follows:
Turn off the apparatus and remove the mains adapter from connector [3-1].
Put the apparatus upside down and remove the two screws fixating the battery compartment cover to
the bottom of the apparatus.
Connect the red lead to the positive (+) pole of the battery and the black lead to the negative (-) pole.
Slide the battery into its compartment.
Reattach the battery compartment cover using the two screws.
Reconnect the mains adapter to connector [3-1].