Manually changing the polarity during a treatment will result in a current decaying to 0, followed by a
current with the opposite polarity, rising to a value equal to 80% of the previous value.
To activate automatic polarity reversal of the The High voltage and Microcurrent waveforms, select
with button [9] and the central controller [16]:
: alternating
: not alternating
With the Interrupted DC current waveform it is not possible to reverse the current polarity.
5.2.1.10 Emergency stop
Pressing buttons [12] and [13] simultaneously will terminate all active treatments.
5.2.2
Electrotherapy
5.2.2.1 2-pole interferential current (premodulated)
Push button [7] to select
channels;
Push button [8], select
Selection of 2-pole interferential current:
Select
with the central controller [16] and press button [14] to accept;
Carrier wave selection:
Push button [9] to select
Beat frequency:
Select
with button [10] and adjust its value with the central controller [16];
Modulation frequency (spectrum):
Select
with button [10] and adjust its value with the central controller [16];
Modulation time (spectrum):
Select
with button [11] and adjust its value with the central controller [16];
Treatment time:
Select
with button [12] and adjust its value with the central controller [16];
Surge program (when desired):
Ramp up time:
Select
with button [12] and adjust its value with the central controller [16];
Hold time:
Select
with button [12] and adjust its value with the central controller [16];
Ramp down time:
Select
with button [12] and adjust its value with the central controller [16];
Interval time:
Select
with button [12] and adjust its value with the central controller [16];
Delay time (only available with linked channels):
Push button [13] and select channel 2 with the central controller [16].
Select
with button [12] and adjust its value with the central controller [16];
Select CC or CV with button [13] and the central controller [16];
Select the desired channel with button [13] and the central controller [16]. Adjust the current intensity
with controller [17];
Push button [14] to start the therapy in case a surge program was selected.
5.2.2.2 Russian stimulation
Push button [7] to select
channels;
Push button [8], select
Carrier wave selection:
Push button [9] to select
Burst frequency:
Select
with button [10] and adjust its value with the central controller [16];
Burst / interval ratio:
Select
with button [11] and adjust its value with the central controller [16];
Treatment time:
Select
with button [12] and adjust its value with the central controller [16];
Surge program (when desired):
Ramp up time:
Select
with button [12] and adjust its value with the central controller [16];
Hold time:
Select
with button [12] and adjust its value with the central controller [16];
Ramp down time:
. Use the central controller [16] to select the desired channel or to link the
with the central controller [16] and press button [14] to accept;
(kHz) and adjust its value with the central controller [16];
. Use the central controller [16] to select the desired channel or to link the
with the central controller [16] and press button [14] to accept;
(kHz) and adjust its value with the central controller [16];
or
GB
7