Resolution
Pressure
Pulse Rate
Measurement Accuracy
Cuff Pressure Accuracy
Error
Operating temperature/ humid-
ity
Transport by general vehicle or according to the order con-
Transport
tract, avoid pounded, shake and splash by rain and snow in
transportation.
Storage
Atmospheric pressure
Power Supply
Battery service life
Rated Power
Dimensions
Unit Weight
Safety classifica-
tion
Service life
Date of manufacturer
Standard Configure:
Adult Cuff: limb circumference 25-35 cm (middle of upper
arm), Software, User Manual, USB data line, BP extending
Accesso-
tube, A pouch.
ries
Separate Sale:
Pediatric Cuff: limb circumference 10-19 cm (middle of upper
arm)
Neonatal Cuff: limb circumference 6-11 cm (middle of upper
neona-
tal
mode
The BP value measured by the device is equivalent with the
measurement value of Stethoscopy, perform clinical verifi-
cation in accordance with the requirements in ISO 81060-2:
2013, whose error meets the followings:
Maximum mean error: ±5 mmHg
Maximum Standard deviation: 8 mmHg
When the temperature is 23 ºC, limb circumfer-
ence is 270 mm, the measured blood pressure is
normal, 2 "AA" alkaline batteries can be used
about 150 times.
Internally powered equipment
Type BF defibrillation-proof applied par
The service life of the device is five years or 10000 times
of BP measurement.
147 ±3 mmHg(19.6 ±0.4 kPa)
1 mmHg(0.133 kPa)
±5 %
Static pressure: ±3 mmHg(±0.4 kPa)
+5 ºC~40 ºC
condensing)
Temperature: -20 ºC~+55 ºC; Relative
humidity: ≤95 %; No corrosive gas and
drafty.
700 hPa~1060 hPa
DC 3 V
≤ 3.0 VA
128(L)*69(W)*36 mm(H)
240 gram(without batteries)
See the label
78
15 %RH~85 %RH(Non-
EN