To insert the rubber bag in the cuff, first place the bag on top of the cuff so
that the rubber tubes line up with the large opening on the long side of the
cuff. Now roll the bag lengthwise and insert it into the opening on the long
side of the cuff. Hold the tubes and the cuff and shake the cuff until the bag
is in position. Thread the rubber tubes from inside the cuff, and out through
the small hole under the internal flap.
Disposable Blood Pressure Cuffs
Disposable cuff is intended for one-patient use only. Do not use the same
cuff on any other patient. No disinfection or high-pressure steam sterilizing
to the disposable cuff. Disposable cuff can be cleaned using soap solution to
prevent infection.
Note
Considering environmental protection, the disposable blood pressure cuffs
must be recycled or disposed properly.
Storage:
Advice
Do not expose the device in direct sunlight for long time, otherwise the dis-
play screen maybe damaged.
The basic performance and safety of the device are not affected by the dust
or cotton wool in home environment, while the device shall not be placed
where are high temperature, humidity, dusty or corrosive gases.
Aged cuff may result in inaccurate measurement, please replace the cuff pe-
riodically according to the user manual.
To avoid device damage, keep the device out the reach of children and pets.
Avoid the device close to extreme high temperature such as fireplace, other-
wise the device performance may be affected.
Do not store the device with chemical medicine or corrosive gas.
Do not place the device where there is water.
Do not place the device where with slope, vibration or impact
4.6 Transportation and Storage
⚫ The packaged device can be transport by general vehicle or according to
the order contract. Do not transport the device mixed with toxic, harmful
or corrosive materials.
⚫ The device after packaged should be stored in a well-ventilated room with-
out any corrosive gas, temperature range: -20℃~+55℃, relative humidity
no more than 95%.