⚫ Using incorrect posture such as standing or lying down, etc.;
⚫ The patient speaks or moves his body during measurement;
⚫ When measuring, the patient is nervous, excited, or in unstable emotion;
⚫ The room temperature rises or falls sharply, or the environment of meas-
urement often changes;
⚫ Measuring in a moving vehicle;
⚫ The position of cuff applied (higher or lower than the heart level);
⚫ Continuous measurement for a long time;
4.4 NIBP Error Messages and Solutions
Display
Cause
message
Low battery
Loose cuff
Atmospheric
pressure error
Signal is
Object measuring the pulse is too
too weak
weak or the cuff is loose.
It is over
Object measuring blood pressure
the range
is over the measurement range.
Excessive
Movement may result in too
move-
much interference in the signal
ment
during measuring process.
Cuff pressure is over the scope,
Over-
ADU 300 mmHg, Newborn:
pressure
150mmHg.
Saturated
Movement or other factors may
signal
lead to too big signal amplitude.
Air leak-
Possible air leakage in the valve
age
or airway
System
Possible failure caused by pump,
failure
air valve or pressure sensor.
The time for a single measure-
It spends
ment exceeds the maximum
too much
measurement time (adult: 180s,
time
neonate: 90s).
Device battery is low.
Cuff is not connected
correctly.
Valve can not be open.
Solution
Replace the battery. If the prob-
lem still exists, please contact us.
Reconnect the cuff. If the prob-
lem still exists, please contact us.
Restart the device. If the problem
still exists, please contact us.
Check the cuff connection,
tighten the cuff if it is loose.
Take another measurement.
If the problem still exists,
please contact us.
Be sure to keep still during
measuring process.
Check the cuff to make sure
it is not blocked or
squeezed.
Check the connection of air
tube to make sure it is not
squeezed. Patient should
keep quiet and then take a
new measurement.
Check the air tube and the
cuff.
Please contact us.
Check the connection of air
tube and tighten the cuff.
63