3.4.1 System Setup
Enter "SYSTEM SET" item in the [SYSTEM MENU], the "SYSTEM SET" menu
includes:
"LANGUAGE" item: switch the current system language;
"UNIT" item has two choices: mmHg, kPa;
"MEASURE MODE" item has three options: adult, pediatric, neonatal;
"ABPM SET" item: set ABPM parameters:
"BACKLIGHT TIME(s)" item: 15, 30, 60, 120
Note
"BACKLIGHT TIME" in the "SYSTEM SET" is used by the common user, ambulatory
blood pressure backlight time is a fixed value of 5s.
To perform ambulatory blood pressure monitoring, first select "ABPM SET"
item in [SYSTEM SET] menu, the pop-up menu is shown in the Figure3.4.2:
Options for "AWAKE INTERVAL (min)" and "ASLEEP INTERVAL (min)": 15, 20,
30, 40, 60, 90, 120, 180, 240;
The step of each adjustment for "AWAKE TIME" and "ASLEEP TIME" is 30
minutes. The adjustment range: 00:00~23:30.
Note
The set measurement interval in "AWAKE INTERVAL" and "ASLEEP INTERVAL"
is time interval when automatically start the measurement under the mode
of ambulatory blood pressure, not including manual starting. For example:
Figure 3.4.1 System Menu
Figure 3.4.2 ABPM Setup
53
EN