RF
emis-
sion
Class B
CISPR
11
Guidance and manufacturer's declaration – electromagnetic immunity –
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of device should assure that it is
used in such an environment.
Immunity
IEC60601
test
test level
Electro-
6 kV con-
static dis-
charge
tact
8 kV air
(ESD)
IEC61000-
4-2
Power
frequency
3A/m
(50/60Hz)
magnetic
field
IEC61000-
4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of device should assure that it is used
in such an environment.
Immun-
IEC60601
ity test
test level
The device is suitable for use in all establishments
other than domestic and those directly connected to
a low voltage power supply network which supplies
buildings used for domestic purposes.
for all EQUIPMENT and SYSTEMS
Compliance
level
6 kV con-
tact
8 kV air
3A/m
Com-
pli-
Electromagnetic environment - guidance
ance
level
Electromagnetic environment
Floors should be wood, con-
crete or ceramic tile. If floor
is covered with synthetic ma-
terial, the relative humidity
should be at least 30%.
Mains power quality should
be that of a typical commer-
cial or hospital environment.