then make corresponding settings according to the following procedure:
"SYS PROMPT" and "DIA PROMPT" can control closing or unsealing of the SYS
and DIA prompt separately.
The prompt is on or off according to the high and low limits which have been
set up. When the measure result is higher than the high limit or lower than
the low limit, and meanwhile, the "PROMPT SOUND" is on, "SYS PROMPT" or
"DIA PROMPT" accordingly on, the prompt will occur.
The adjustable ranges of the high and low limits of the adult mode prompt
are as follows:
SYS PROMPT: 40~270 mmHg
DIA PROMPT: 10~215 mmHg
The adjustable ranges of the high and low limits of the pediatric mode
prompt are as follows:
SYS PROMPT: 40~200 mmHg
DIA PROMPT: 10~150 mmHg
The adjustable ranges of the high and low limits of the neonatal mode
prompt are as follows:
SYS PROMPT: 40~135 mmHg
DIA PROMPT: 10~100 mmHg
"DEFAULT" includes the main content:
Measure mode: adult;
Parameter prompting limit:
User
Systolic pres-
mode
sure high limit
Adult
Pediatric
Neonatal
PROMPT SOUND switch: OFF;
Measure unit: mmHg;
Ordinary user backlight time: 120s;
ABPM switch: END;
Asleep time: 22:00;
Asleep measurement interval: 30minutes;
Awake measurement interval: 15minutes;
Awake time: 7:00;
SYS PROMPT switch: OFF;
DIA PROMPT switch: OFF.
Note: The monitor does not have an alarm system.
3.4.5 ABPM Menu
1.ABPM mode
After the ambulatory blood pressure menu is operated correctly (refer to
3.4.1), select "ABPM" menu in [SYSTEM MENU] to enter its interface.
Systolic pres-
sure low limit
140
120
90
Diastolic pres-
sure high limit
90
70
60
55
Diastolic pres-
sure low limit
90
70
60
EN
40
40
20