the report, then select "Print" to print the report.
6.19 Help
Click the shortcut key
for each program function. In addition, you will find "Help" button in each
operation interface, click it to check the description for this function, which
is convenient for you to quick know the use of software.
Specification
Name
The
degree
against ingress of water
Display
Operating mode
NIBP Specifications
Measurement Method
Working modes
Cuff pressure range
Blood pres-
sure meas-
pediatric
urement
Range
neonatal
Pulse
measurement
range
Inflation
Prompt
pediatric
Range
neonatal
Overpressure protect
to its sub-menu, which gives a brief description
of
protection
Automatic
adult
pediatric 0~235 mmHg(0~31.3 kPa)
neonatal 0~147 mmHg(0~19.6 kPa)
SYS:40~270 mmHg(5.3~36.0 kPa)
adult
DIA:10~215 mmHg(1.3~28.7 kPa)
SYS:40~200 mmHg(5.3~26.7 kPa)
DIA:10~150 mmHg(1.3~20.0 kPa)
SYS:40~135 mmHg(5.3~18.0 kPa)
DIA:10~100 mmHg(1.3~13.3 kPa)
40~240/min
adult
pediat-
ric
neona-
tal
adult
SYS PROMPT: 40~270 mmHg(5.3~36.0 kPa)
mode
DIA PROMPT: 10~215 mmHg(1.3~28.7 kPa)
SYS PROMPT: 40~200 mmHg(5.3~26.7 kPa)
mode
DIA PROMPT: 10~150 mmHg(1.3~20.0 kPa)
SYS PROMPT: 40~135 mmHg(5.3~18.0 kPa)
mode
DIA PROMPT: 10~100 mmHg(1.3~13.3 kPa)
adult
mode
pediat-
ric
mode
Ambulatory Blood Pressure Monitor
IP22
2.4'' color LCD Display
Continuous operating
Oscillometric method
0~297 mmHg(0~39.6 kPa)
160mmHg(21.33 kPa)
120mmHg(16kPa)
70mmHg(9.33 kPa)
297±3 mmHg(39.6±0.4 kPa)
240±5 mmHg(32±0.67kPa)
77