damage around the catheter when infusion is slowed or blocked during cuff
inflation.
The minimum value of the patient's physiological signal is the lower limit that
the device can measure. The measured result may inaccurate if the device
running below the minimum amplitude or minimum value of patient's phys-
iological signal.
Do not twist or tangle the airway tube, otherwise it will cause continuous
pressure in the cuff; then causing blocked blood flow and serious injury to
the patient.
Do not use the cuff on the injured area; otherwise, it will cause more serious
damage to the injured area.
Do not use the cuff on the site where intravascular treatment is being per-
formed or with catheter connection, otherwise it may cause temporary
blockage of blood flow and then cause injury to the patient.
Do not use the cuff on the side of the mastectomy;
The pressure by cuff may cause temporary weakness of some functions of
the body. So do not use monitoring medical electrical equipment on corre-
sponding arm.
Do not move during measuring, because it will have a delayed effect on the
patient's blood flow.
The device needs 2 hours recovery time to reach its performance of intended
use after taking out from the lowest storage temperature.
The device needs 4 hours recovery time to reach its performance of intended
use after taking out from the highest storage temperature.
1. Plug the air hose to the cuff socket on the device, and connect the device
to power supply.
2. Apply the cuff to the patient's upper arm following the instructions below
(Figure4.2.1).
⚫ Ensure the cuff is completely deflated.
⚫ Apply the appropriate size cuff to the patient, and make sure that the
symbol "φ" is over the appropriate artery. Ensure that the cuff is not wrapped
too tightly around the limb. Excessive tightness may cause discoloration and
eventual ischemia of the extremities.
Figure 4.2.1 Cuff applying
60