Chapter 4

unpacking, assembly and system Check

4.1

Overview

This section describes how to unpack the pump and connect the accessories . After unpacking
and assembly, follow the system check procedure to ensure that the FMS VUE II System is
working properly .
4.2

unpacking

to unpack the pump:
1 . Place hands in the open areas between the two pieces of packaging foam on both sides of
the console until you can securely grasp the underside of the pump .
2 . Carefully remove the pump from the box and place it on a level and sturdy surface .
3 . Verify that all items on the packing list match the products you have received .
4 . Report damage immediately to a DePuy Synthes Sales Representative or call
DePuy Synthes Customer Service in the USA at +1-800-382-4682 . Save all packing
materials to use in the event you must return product(s) to DePuy Synthes . Outside the
United States, contact your local affiliate .
5 . Recycle all FMS VUE II packaging material, including cardboard, foam inserts and anti-
static bag .
4.3

pump placement

Position the FMS VUE II pump outside the sterile field, and where required depending on
operator requirements . Ensure that there is enough space for Circulating Nurses to access the
pump controls . Position the pump so that access to power main is not obstructed .
There are two options for setting up pump elevation:
• Place the pump (fill line of Fill Chamber) at the same level as the patient joint .
• Place the pump at a convenient cart elevation and adjust the joint-to-pump elevation from the
front panel at startup . Refer to Section 6 .1 .1, "Setting Joint-to-Pump Elevation" .