nOte: The elevation factory default setting of 0 cm can be modified via the User Default menu.
(Section 6.2).
if the joint is at the same level as the fill Chamber:
Verify/Adjust the elevation setting to 0 cm using the "+/-" pads on the right display .
if joint is above pump fill Chamber:
1 . Estimate or measure the distance between the joint and the pump Fill Chamber
(e .g .+90 cm)
2 . Verify/Adjust the elevation setting as required using the "+/-" pads on the right display .
if joint is below pump fill Chamber:
1 . Estimate or measure the distance between the joint and the pump (e .g .-60 cm)
2 . Verify/Adjust the elevation setting as required using the "+/-" pads on the right display .
3 . Press the run/stOp pad to select set-point for the desired joint pressure .
32
Joint-to-Pump
Elevation (AT)
Joint Elevation
Fill Chamber
(BELOW)
Fill Chamber
(ABOVE)
Joint Elevation