Characteristics; Composition; Ingress Protection; Dimensions - Johnson & Johnson DePuy Synthes FMS VUE II Manuel De L'opérateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 293
Chapter 10
Specifications and Performance

Characteristics

10.1 Composition

Ref: 284004 FMS VUE II Pump

10.2 ingress protection

Ingress Protection Rating: IP40

10.3 dimensions

Height: 21 cm
Width: 38 .5 cm
Depth: 37 cm
Weight: 12 .8 kg

10.4 electrical specifications

input voltage: 100-240 V
frequency: 50/60 Hz
input power: 500 VA
Fuse ratings: 5 Amps, 250 Volts, Breaking Capacity: H
Conforms to standard IEC 60601-1 Medical Device Electrical Safety and IEC 60601-1-2
(Electromagnetic Compatibility)

10.5 degree of protection (applied parts)

The FMS VUE II pump is a Class 1 device according to IEC 60601-1 . (No electrical contact with
patient) . BF isolation applies to both pumps and shaver .

10.6 performance Characteristics

Joint pressure range: 20–150 mmHg (increments of 5) . When pressure is stabilized, levels
will be within 20% of the set-point .
Cannula flow rate: 100–600 ml/min .
Pre-selected Precision (±20%) of maximum flow rate .
handpiece flow rate: 200–800 ml/min .
Pre-selected Precision (±20%) of maximum flow rate .
59

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

284004

Table des Matières