Chapter 10
Specifications and Performance
Characteristics
10.1 Composition
Ref: 284004 FMS VUE II Pump
10.2 ingress protection
Ingress Protection Rating: IP40
10.3 dimensions
Height: 21 cm
Width: 38 .5 cm
Depth: 37 cm
Weight: 12 .8 kg
10.4 electrical specifications
input voltage: 100-240 V
frequency: 50/60 Hz
input power: 500 VA
Fuse ratings: 5 Amps, 250 Volts, Breaking Capacity: H
Conforms to standard IEC 60601-1 Medical Device Electrical Safety and IEC 60601-1-2
(Electromagnetic Compatibility)
10.5 degree of protection (applied parts)
The FMS VUE II pump is a Class 1 device according to IEC 60601-1 . (No electrical contact with
patient) . BF isolation applies to both pumps and shaver .
10.6 performance Characteristics
Joint pressure range: 20–150 mmHg (increments of 5) . When pressure is stabilized, levels
will be within 20% of the set-point .
Cannula flow rate: 100–600 ml/min .
Pre-selected Precision (±20%) of maximum flow rate .
handpiece flow rate: 200–800 ml/min .
Pre-selected Precision (±20%) of maximum flow rate .
59