guidance and manufaCturer's declaration – immunity
guidance and manufacturer's declaration – emissions
The FMS VUE II is intended for use in the electromagnetic environment specified below . The
customer or user of the FMS VUE II should ensure that it is used in such an environment .
immunity test
ESD
+/-8kV Contact
IEC 61000-4-2
+/-15kV Air
EST
+/-2kV Mains
IEC 61000-4-4
+/-1kV I/Os
100KHz frequency
repetition
Surge
+/-1kV Line-to-line
IEC 61000-4-5
+/-2kV Line-to-
ground
Voltage Dips/
100% Dip for
Dropout
0 .5 Cycle
IEC 61000-4-11
Single phase at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315°
100% Dip for 1
cycle and 30% Dip
for:
- 25 Cycles at
50Hz
- 30 Cycles at
60Hz
100% Dip for:
- 250 Cycles at
50Hz
- 300 Cycles at
60Hz
Power Frequency
30 A/m
50/60Hz
Magnetic Field
IEC 61000-4-8
all me equipment and me systems
ieC 60601
Compliance level
test level
+/-8kV Contact
+/-15kV Air
+/-2kV Mains
+/-1kV I/Os
100KHz frequency
repetition
+/-1kV Line-to-line
+/-2kV Line-to-
ground
100% Dip for
0 .5 Cycle
Single phase at 0°,
45°, 90°, 135°, 180°,
225°, 270°, 315°
100% Dip for 1 cycle
and 30% Dip for:
- 25 Cycles at 50Hz
- 30 Cycles at 60Hz
100% Dip for:
- 250 Cycles at 50Hz
- 300 Cycles at
60Hzs
30 A/m
70
electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile . If
floors are synthetic, the r/h
should be at least 30% .
Mains power quality should be
that of a typical commercial or
hospital environment .
Mains power quality should be
that of a typical commercial or
hospital environment .
Mains power quality should be
that of a typical commercial or
hospital environment .
If the user of the FMS VUE II
requires continued operation
during power mains
interruptions, it is recommended
that the FMS VUE II be
powered from an uninterruptible
power supply or battery .
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment .