Johnson & Johnson DePuy Synthes FMS VUE II Manuel De L'opérateur page 71

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 293
guidance and manufaCturer's declaration – immunity
me equipment and me systems that are nOt life-supporting
guidance and manufacturer's declaration – emissions
The FMS VUE II is intended for use in the electromagnetic environment specified below . The
customer or user of the FMS VUE II should ensure that it is used in such an environment .
immunity test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
6Vrms in ISM bands
between 150KHz and
80MHz
80% AM at 1KHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2 .7 GHz
80% AM at 1KHz
* Vrms = Voltage Root Mean Square - DC equivalent voltage
ieC 60601
Compliance level
test level
(V1) = 3 Vrms
(V2) = 6 Vrms
(E1) = 3 V/m
71
electromagnetic
environment - guidance
Portable and mobile
communications equipment
should be separated from
the FMS VUE II by no
less than the distances
calculated/listed below:
D=(3.5/V1)(√ P)
150kHz to 80MHz
D=(3.5/V2)(√ P)
in ISM bands between
150KHz to 80MHz
D=(3.5/E1)(√ P)
80 to 800 MHz
D=(7/E1)(√ P)
800 MHz to 2 .5 GHz
Where P is the max power
in watts and D is the
recommended separation
distance in meters .
Field strengths from fixed
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance levels (V1
and E1) .
Interference may occur in
the vicinity of equipment
containing a transmitter .