fault Code
FC004
Pressure sensor/electronics
problem detected
FC005
Shaver motor over-current fault
FC006
Shaver motor controller fault,
caused by thermal fault, or over-
current fault
FC007
Hand-piece 5V over-current fault
FC008
Foot-pedal 5V over-current fault
FC009
Irrigation motor over-current fault
fault description
57
fault Correction
Pressure sensors readings difference
detected .
Discontinue use of the console . Contact
your DePuy Synthes representative .
Possible short circuit in shaver, replace
shaver to avoid possible damage to the
console .
Possible short in shaver or damage to
motor controller .
Replace shaver . If replacing shaver
doesn't correct issue, contact your
DePuy Synthes representative .
Possible short in hand-piece connector
or cable, replace hand-piece .
Possible short circuit in foot-pedal
connector or cable, replace foot-pedal
Possible pump motor short circuit .
Discontinue use of the console . Contact
your DePuy Synthes representative .