OVERVIEW

4
The Dentsply Sirona HVE Mirror Tip and Hose Adapter Starter Kit offers a new and unique dental solution for high volume
evacuation (HVE). The kit includes three (3) HVE Mirror Tips with an ergonomic unibody design and built-in mirror that allows
4
for evacuation, illumination, retraction, and visualization with one hand. The HVE Hose Adapter allows the HVE Mirror Tip to be
4
used in the most effective and comfortable manner, with a lightweight, flexible, and durable hose that provides greatly improved
ergonomic support. The HVE Hose Adapter connects directly to a standard 11 mm HVE valve and also includes a 360° swivel
4
connection for the HVE Mirror Tip and a holster to secure the system when not in use.
5
5

INTENDED USE

6
The Dentsply Sirona HVE Mirror Tip and Hose Adapter Starter Kit is intended to provide oral high volume evacuation, visualization,
illumination, and retraction during dental procedures. The device is intended to be used by trained dental professionals only.
6
6

CONTRAINDICATIONS

7-9
There are no known contraindications.
10-13
14

WARNINGS

14
• Only use compatible HVE tips with the Dentsply Sirona HVE Hose Adapter.
• Sterilizing the HVE Hose Adapter will cause component damage. Disinfect the HVE Hose Adapter using only the tested and
approved disinfectants listed in the Infection Control Procedures Section.
• The HVE Mirror Tips must be steam sterilized before first use and between patients to prevent patient cross-contamination. See
Infection Control Procedures Section.
• The HVE Mirror Tip is designed to last up to 100 reprocessing cycles and the HVE Hose Adapter is designed to last up to 2,500
reprocessing cycles. Use of the device beyond its useful life may cause damage to equipment and increases risk of patient cross-
contamination.