17. Disposal of Unit

U.S. -Dispose of the system components in accordance with state and local laws.
EU - Dispose of in accordance with the Waste Electrical and Electronic Equipment Directive 2002/96/EC of the
European Parliment and the Council of the European Union.
18. Electro-Magnetic Compatibility Precautions

Guidance and manufacturer's declaration - electromagnetic emissions

The Nupro Freedom® is intended for use in the electromagnetic environment specified below. The Customer or the
user of the Nupro Freedom® should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Nupro Freedom® is intended for use in the electromagnetic environment specified below. The customer or the
end user of the Nupro Freedom® should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
±6 kV contact
Electromagnetic
±8 kV air
discharge (ESD)
IEC 61000-4-2
±2 kV for power
Electrical fast
transient/burst
supply lines
±1 kV for input/out-
IEC 61000-4-4
put lines
±1 kV line(s) to
Surge
IEC 61000-4-5
line(s)
±2 kV line(s) to
earth
Voltage dips,
<5 % U
short interrup-
(>95 % dip in U
tions and volt-
for 0.5 cycle
age variations
40 % U
on power supply
(60 % dip in U
input lines
for 5 cycles
IEC 61000-4-11
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
English • 18
Compliance
Group 1
Class B
Class B
Complies
Compliance level
±2, ±4, ±6kV contact
±2, ±4, ±8kV air
±0.5, ±1, ±2kV mains
No I/O lines
±0.5, ±1kV Line to Line
±0.5, ±1, ±2kV Line to
Earth
<5 % U
T
T
)
(>95 % dip in U
T
for 0.5 cycle
40 % U
T
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
T
)
(>95 % dip in U
T
for 5 sec
3 A/m
Electromagnetic environment - guidance
The Nupro Freedom® uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The Nupro Freedom® is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Warning: This equipment/system is intended for use by
healthcare professionals only. This equipment/system may
cause radio interference or may disrupt the operation of
nearby equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the Nupro
Freedom® or shielding the location.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typi-
)
cal commercial or hospital environment. If the
T
user of the Nupro Freedom® requires continued
operation during power mains interruptions, it
is recommended that the Nupro Freedom® be
)
powered from an uninterruptible power supply or
T
a battery.
)
T
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typi-
cal commercial or hospital environment.