Appendix B: System Components And Accessories; Ipgs - St.Jude Medical Proclaim Manuel

Générateur d'impulsions implantable (gii)
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Model
Connector
strength
Program storage
capacity
* Battery longevity was estimated using the following nominal settings 12 hours per day: 50-Hz
frequency, 225-μs pulse width, and 5-mA amplitude at 500-ohms impedance. For information on
how additional settings may impact the longevity of the device, please contact Technical Support.
The IPG has the following operating parameters.
Table 4. Operating parameters for the IPG
Parameter
Tonic Range
Pulse width
20–1000 µs
Frequency
2–200 Hz
200–500 Hz
500–1200 Hz
Burst rate
frequency
Intraburst
frequency
Amplitude
0–25.5 mA
NOTE: The number of stim sets in use for a tonic program governs the maximum frequency
(1200/number of stim sets).
NOTE: The maximum current depends on the impedance, frequency, and pulse width settings.

Appendix B: System Components and Accessories

The Proclaim™ neurostimulation system includes the following components.
NOTE: Not all models are available in all countries. Contact your local representative for more
information.

IPGs

3660 Proclaim™ 5 Elite IPG
3662 Proclaim™ 7 Elite IPG
3665 Proclaim™ 5 IPG
3667 Proclaim™ 7 IPG
ARTMT100145789A.pdf 14
Exceeds EN 45502-1 requirements
16 programs with 8 stim sets each
Steps
10 µs
(20–500 µs range)
50 µs
(500–1000 µs
range)
2 Hz
10 Hz
20 Hz
0.1–1.0 mA
12
MRI Status
Upgradeable
Features
Burst Range
Steps
20–1000 µs
50 µs
10–60 Hz
10 Hz
250–500 Hz
10 Hz
500–1000 Hz
20 Hz
0–12.75 mA
0.05–1.0 mA
11/11/2015 8:52:58 AM