Medical Device Directive (Europe)
This device complies with the requirements of Directive
93/42 EEC concerning medical devices
0123
Use
This user manual is valid for the products listed on page 95.
HF cables connect electrode handles, electrodes and in-
struments for monopolar or bipolar coagulation to HF units.
When used as intended, these products are more or less
subject to wear, depending on the intensity of use. This
kind of wear is caused by technical factors and cannot be
avoided. Therefore carry out a thorough visual examination
and a function test prior to each application of these
products. Only use compatible products for plug
connections and make sure all the plugs are inserted
properly. A product must be replaced if it shows signs of
wear or fails to function as described in these instructions.
In this case, please contact the manufacturer or one of its
authorized representatives. Also see Chapter 4. Repairs.
Operating
Thoroughly examine all products for visible irregularities.
Do not use cables with a brittle or defective insulation. In
case of doubt check operation by means of a suitable
testing device (e.g. 050-230).
Generators with Plug'n Cut COMFORT can recognise
BOWA COMFORT instruments and automatically select the
appropriate parameters.